The current status of the 8pi Germanium detectors is shown below. (last updated 03 February 2012)

Det# Serial# Status BIAS (V) Last Factory visit Last Pumped Last Tested FWHM @ 1333 keV Comment
01 25-N1406A -Lab -3000 1999 2008 Feb 13/Feb/08 Spiky under biasing
02 27-N07LA M -2000 2004 2008 Aug 26/Aug/08 Good detector
03 25-N1365B P -4000 1988 2012 Mar 03/Apr/12 2.15 Good detector
04 25-N1481A G -3000 2011 2011 Jul Requires special injector and mounting plate.
05 25-N-1454 X -3000 2008 2010 Aug 13/Aug/10 Good Detector
06 25-N-05RB -Lab -3000 2002 2009 Oct 26/Oct/09 Good detector but preamp signal intermittant
07 25N1415A N -2000 1986 2012 Feb 13/Feb/12 2.07 Good detector
08 27NQ50 S -3000 1995 2008 Jul 29/Aug/08 Good detector
09 24-N87SA W -3500 2004 2010 Nov 09/Nov/10 Good detector
10 25-N1260B T -2800 2002 2009 Jan 06/04/05 Good detector
11 25-N18QB lab -3000 1997 2008 Jan 01/Feb/08 pumping station
12 25-N1494B K -2000 2009 2009 Factory 02/Mar/09 Good detector
13 25-N1446 C -3000 2002 2012 Feb 09/Mar/12 1.96 Good detector
14 24-N09RB H -3000 2008 May 01/Jun/08 Good detector
15 27-NQ41 R -2500 2004 2010 Feb 01/Mar/10 Little spiky under biasing
16 27-NQ03B B -2500 1998 2009 Oct 15/Oct/09 Good detector
17 25-N1384B -Lab -2500 2004 2010 Aug 07/Nov/10 Reported good in Nov 2010
18 25-N1370B F -3000 2004 2009 Jan 02/Feb/09 Good detector
19 25-N50RB J -2000 2002 2010 Oct 17/Oct/10 Good detector
20 24-N905B D -2500 2004 2009 Dec 24/May11 Good Detector
21 24-N96SB E -4000 2010 2010 Factory 21/July/10 Good Detector
22 25-N1452 Y -3000 2009 2011 Jun 14/Jun/11 Good Detector
23 25N1412B -Lab -2500 2011 2011 Factory 07/Feb/11 Unstable baseline
24 27-N1952B L -3000 2010 2012 Apr 18/Apr/12 1.91 Good detector
25 25-N21Q A -2000 2008 2008 Factory 21/Oct/09 Bad - Strange risetime.