TIGRESS Clover Initial Testing

From tigwiki

Jump to: navigation, search

Contents

[edit] Initial Testing of Clovers

[edit] Biasing of Crystals

  • There is a set of HV cables for biasing. Connect the shut down and the HV cables between the clover and the HV supply CAEN SY2527 tighv03
  • Signal cables from the preamps are colour coded. (Blue, Green, Red and White)
  • Full volume energy signals #10 (FV) are connected from each crystal over to the scope.
  • To control bias
    • ssh midtig03, account tigress, normal password
    • telnet tighv03 1527
  • Move around shell with tab, arrows.
  • The voltages for all four crystals are arranged as a group so that four voltages are increased together.
  • Bias is increased at 3 volts/sec. Step up in 100 Volt steps to 500V and then in 500V steps up to recommended operating voltage (written on clover dewar)
  • After a few hundred volts inspect the signals to see that (negative) rise time is of order 50-100 ns and fall time ~ 50 us
  • If sparking occurs , lower voltage below the sparking voltage and let sit. (condition)
  • After full bias is reached, add sources and let sit for a day.

[edit] Preamp gain

  • Preamp output should be 200 mV/ MeV.
  • With 50 ohm input impedance on scope,then 137Cs line will be about 65 mV.

[edit] Resolution testing

  • use 60Co and 241Am sources
  • Each of the channels is then fed, in turn, through amplifier to MCA.
  • Shaping time 4 us (Canberra uses 6us but 4us appears to give better resolution with this amplifier).
  • Adjust the pole-zero.
  • calibrate spectrum.
  • for FV signals accumulate ~10,000 counts.
  • for sectors about 1,600 with sources somewhat closer.
  • for back sector must move Am241 around (won’t be seen for sector 7)
  • store spectra.

[edit] Preamp Risetime Measurement

It is not necessary to make the four incorrect measurements that I did so they won’t be listed here.

  • Feed pulses from HP 8082A pulse generator to test input of clover.
  • 1 kHz rate and 1 ns transition time.
  • FV output through short cable to scope. Trigger scope from pulse generator.
  • select measure on scope to set an automatic fall time measurement.

[edit] Efficiency measurement

Standard Co60 source is placed 25 cm from front of Clover. (There is a modified ruler which is 25 cm long.)

  • Accumulate spectra for all four crystals noting live times.
  • Calculate the yield for the 1332 keV line in c/s for each crystal.
  • Source used was 38.48 kBq on June 1, 2007. Determine time since then in days.
  • Half life is 5.2714 years.
  • Therefore source strength is now 38480 exp (0.693 * days/365.25/ 5.2714).

Germanium detector efficiencies are measured relative to that of a 3" x 3" NaI crystal which has a relative efficiency of 0.0012 at 25 cm for 1332 keV.

Then crystal efficiency expressed as a percentage = yield / source strength / 0.12 .